The Total Dissolved Gas Instruments tests for the following:

Manuals

T300E Manual
T506 Manual
T507 Manual
T900 Manual

Additional Resources

T300E Gas Calculation (Excel Spreadsheet)
T300E Gas Calculation (PDF)

Total Dissolved Gas Meters

Total Dissolved Gas - TDG Sensor
 
 

Supersaturated water can kill your fish

Dissolved oxygen measurements are not enough--
oxygen levels may be acceptable while nitrogen levels may be lethal

Get the entire picture -- the sum total pressure of all dissolved gases

Calculate NITROGEN levels from TDG and a dissolved oxygen measurement

Monitor for supersaturated waters with one unit!

Total dissolved gas pressure is an important parameter for monitoring water quality effectively. When the total pressure of all the gases in the water exceeds the ambient atmospheric pressure at the water surface, supersaturation exists. The effect of excessive supersaturation on fish has been well documented, and if supersaturation exceeds the established safe levels, massive fish kills can occur.

Supersaturation can be caused by numerous man made and natural sources:

  • Dam spillways which allow the discharge to plunge deeply into the receiving water
  • Heated discharge of thermo-electric power plants can raise the temperature of the receiving water to a point of supersaturation
  • Faulty water pumps used in aquaculture recirculatory systems can cause air entrainment if a leak occurs on the suction side of the pump
  • Naturally high levels of nitrogen can be found in well water
  • Algae blooms can cause increased levels of gas in the water

The T300E SensorPrecautions can be taken to keep supersaturation within safe levels, but accurate in-situ gas measurements must be made to determine if remedial action is effective. The TDG Sensor makes this measurement quick and easy

Applications:

  • Use to monitor fish hatcheries, rearing ponds, holding tanks, aquariums, dams, etc. wherever fish are present
  • Check well water and other critical water sources for age or for supersaturation caused by heated water or faulty pumps
  • Check for supersaturation caused by heated water discharge at power plants, industrial sites, and sewer outfalls

Versions
TDG is available in multiple versions to meet the needs of water professionals:

  T506 T507 T900-4 T900-10 T900-20
Measurement type Absolute Absolute Dirrerential Differential Differential Pseudo differential against electrical signal. The sensor is an absolute type. Barometric is cancelled electrically.
Equipment Probe Probe Probe/cable Probe/cable Probe/cable Meter/Probe/cable
Measurement Method Membrane Diffusion Membrane Diffusion Membrane Diffusion Membrane Diffusion Membrane Diffusion Membrane Diffusion

Measurement Performance

Pressure Range/Output Units 2 atm 400 to 1400 mmHg +/- 400 mmHg +/- 400 mmHg +/- 400 mmHg +/- 750 mmHg
Pressure Resolution 0.0001 Atm 0.5 mmHg 0.5 mmHg 0.5 mmHg 0.5 mmHg 1 mmHg
Pressure Accuracy +/- 0.002 Atm +/- 1.5 mmHg +/- 1.5 mmHg +/- 1.5 mmHg +/- 1.5 mmHg +/- 1 mmHg
Equilibration Time to 90% of final 5 min 5 min 5 min 5 min 5 min 5 min
Output Temp Range -30 to +50 ° C -30 to +50 ° C n/a n/a n/a n/a
Output Temp Resolution .0.1 ° C +/- 0.25 ° C n/a n/a n/a n/a
Output Temp Accuracy +/- 0.25 ° C +/- 0.25 ° C n/a n/a n/a n/a
Output
Voltage n/a 0.5 to 5.0 volt 0 to 5.0 volt 0 to 5.0 volt 0 to 5.0 volt n/a
Digital ASCII : 9600,8,N,1 ASCII : 9600,8,N,1 ASCII : 9600,8,N,1 ASCII : 9600,8,N,1 ASCII : 9600,8,N,1 3 Digit LCD
Signal Output digital digital/analog digital/analog digital/analog digital/analog digital (LCD display only)

Physical

Water matrix Natural waters, either fresh, brackish, or saline Natural waters, either fresh, brackish, or saline Natural waters, either fresh, brackish, or saline Natural waters, either fresh, brackish, or saline Natural waters, either fresh, brackish, or saline Natural waters, either fresh, brackish, or saline
Material Delrin, silicone, stainless steel Delrin, silicone, stainless steel Delrin, silicone, stainless steel Delrin, silicone, stainless steel Delrin, silicone, stainless steel Delrin, PVC, silicone, stainless steel
Probe Dimensions 300mm x 45mm 300mm x 45mm 300mm x 45mm 300mm x 45mm 300mm x 45mm 127mm x 29mm
Weight n/a n/a n/a n/a n/a 0.9 kG
Membrane Cartridge P/N 70-020 P/N 70-020 P/N 70-020 P/N 70-020 P/N 70-020 P/N 70-020
Deployment time before cleaning 2-4 weeks 2-4 weeks 2-4 weeks 2-4 weeks 2-4 weeks not for long term deployment. Spot checker
Electrical
Supply Voltage 6.0 to 13.0 Volts DC External 8.6 to 13.6 Volts DC External 8.0 to 16.0 Volts DC External 8.0 to 16.0 Volts DC External 8.0 to 16.0 Volts DC External 2 Internal 9V alkaline batteries
Power consumption (continuous). Duty cycle dependent on controller. 7 mA 7 mA 7 mA 7 mA 7 mA 1.5 mA

Enviornmental

Storage Temp w/ membrane 0 °C to +60 °C 0 °C to +60 °C 0 °C to +60 °C 0 °C to +60 °C 0 °C to +60 °C 0 °C to +60 °C
Storage Temp w/o membrane -40 °C to +60 °C -40 °C to +60 °C -40 °C to +60 °C -40 °C to +60 °C -40 °C to +60 °C -40 °C to +60 °C
Operating Temp 0 °C to +45 °C 0 °C to +45 °C 0 °C to +45 °C 0 °C to +45 °C 0 °C to +45 °C 0 °C to +45 °C
Operational Depth < 53 meters < 53 meters < 53 meters < 53 meters < 53 meters < 53 meters
Max Depth 60 meters 60 meters 60 meters 60 meters 60 meters 60 meters

Accessories

Battery/PC connector available available n/a n/a n/a n/a
Cable Available but not included in list price. Waterproof oceanographic plug Available but not included in list price. Waterproof oceanographic plug 4 meters fixed included. Vented. Bare wire end for connection to controller 10 meters fixed included. Vented. Bare wire end for connection to controller 20 meters fixed included. Vented. Bare wire end for connection to controller 5 meters fixed included. Vented. Bare wire end for connection to controller
Pressure Cap INCULDED INCELUDE OPTIONAL OPTIONAL OPTIONAL OPTIONAL
Membrane Cartridge One Inculded One Inculded One Inculded One Inculded One Inculded One Inculded
Application Aquifer testing. Aging of water, remediation air sparging, cavitations Hydroelectric, fish protection, cavitations Aquaculture, Fisheries, Commercial Aquariums Aquaculture, Fisheries, Commercial Aquariums Aquaculture, Fisheries, Commercial Aquariums Aquaculture, Fisheries, Commercial Aquariums

T506 T507 T900
T300E
General
Description Sensor only Sensor only Sensor with 4, 10,
or 20m cable
Sensor with 10m
cable w/meter
Unit Type Digital Digital/Analog Digital/Analog Digital (LCD display only)
Material Delrin, silicone,
stainless steel
Delrin, silicone,
stainless steel
Delrin, silicone,
stainless steel
Delrin, PVC, silicone, stainless steel
Media Compatibility Natural waters, either fresh, brackish, or saline Natural waters, either fresh, brackish, or saline Natural waters, either fresh, brackish, or saline Natural waters, either fresh, brackish, or saline
Cable connection to Probe Detachable non-vented oceangraphic connector Detachable non-vented oceangraphic connector Fixed vented cable Fixed non-vented w/connector on the
meter end
Probe Dimension 300mm X 45mm 300mm X 45mm 300mm X 45mm 127mm X 29mm
Weight 0.9 kG
Handheld option -- -- RuggedReader,
Pocket PC, PalmPilot
Analog Meter with
LCD display
 
Environmental
Storage Temp w membrane 0C to 60C 0C to 60C 0C to 60C 0C to 60C
Storage Temp wo membrane -40C to 60C -40C to 60C -40C to 60C -40C to 60C
Operating Temp 0C to 45C 0C to 45C 0C to 45C 0C to 45C
Operational Depth 53 meters 53 meters 53 meters 53 meters
Max Depth 60 meters 60 meters 60 meters 60 meters
 
Power
Power Supply External 6-13 VDC External 8.6-13.6 VDC External 8-16 VDC 2 Internal 9V alkaline batteries
Power Consumption 7 mA 7 mA 7 mA 1.5 mA
 

Output

Voltage n/a 0.5 to 4.5 volt 0 to 5.0 volt n/a
Digital ASCII : 9600,8,N,1 ASCII : 9600,8,N,1 ASCII : 9600,8,N,1 3 DIGIT LCD
 

TDG Sensor

Type Non-vented Non-vented Vented Non-vented (Barometric reading can be zeroed)
Membrane cartridge P/N 70-020 P/N 70-020 P/N 70-020 P/N 70-020
Recommended Deployment Interval 2-4 weeks 2-4 weeks 2-4 weeks 2-4 weeks
Measurement Method Membrane diffusion Membrane diffusion Membrane diffusion Membrane diffusion
Pressure Range 2 Atm 400 to 1400 mmHg +/- 400 mmHg +/- 750 mmHg
Pressure Resolution 0.0001 Atm 0.5 mmHg 0.5 mmHg 1 mmHg
Pressure Accuracy +/-0.002 Atm +/- 1.5 mmHg +/- 1.5 mmHg +/- 1 mmHg
Eqilibration Time to
90% of final
5 min 5 min 5 min 5 min
Output Temp Range -30 to +50 deg C -30 to +50 deg C -- --
Output Temp Resolution 0.1 deg C 0.1 deg C -- --
Output Temp Accuracy +/- 0.25 deg C +/- 0.25 deg C -- --
 

Accessories

PC Connection Available Available -- --
Cable Connections Available but not included in list price. Waterproof oceanographic plug Available but not  included in list price. Waterproof oceanographic plug Included. Choice of three fixed lengths. Vented. Bare wire end for connection to controller. 5 meter length included. Custom lengths available.
Pressure Cap Included Included Extra Extra
Membrane Cartridge One included One included One included One included

Request additional information | Request price quote

Web Effects (c) Copyright, 2005 OpenCube Inc. All Rights Reserved.